Publicação:
Wear of the implant connection with zirconia and titanium abutments

dc.contributor.authorJosé, A. R.
dc.contributor.authorMarco, M.
dc.contributor.authorMaurício, P. J.
dc.contributor.authorDiogo, M.
dc.contributor.authorFátima, Vaz M.
dc.contributor.authorLuís, R.
dc.contributor.authorIgnacio, B. N.
dc.date.accessioned2019-07-23T15:49:18Z
dc.date.available2019-07-23T15:49:18Z
dc.date.issued2017-10
dc.description.abstractBackground: Classify and quantify the wear of the implant and abutment interface simulating a loose screw situation. Methods: Nine external connection titanium implants were split into 4 groups: Group A (no hexagon titanium abutment), Group B Titanium abutment (with hexagon) (n=3), Group C Zirconia abutment with hexagon (n=3) and Group D mix abutments (n=2). All components were individually weighted and the rotational freedom for each pair was access before and after testing. The specimens were tested an Instron 8874 fitted with a dynamic load cell with a torque capacity of 100Nm and a precision of 0.5% from 1% of the full scale. They were loaded with a sinusoidal rotational angle that was different for each group (A= 6º, B= 3.3º, C= 0.835º, D=6.6 º), 4 Hz, for 250.000 cycles. After testing all components were photographed with a SEM. Spearman’s correlation analysis was made. Results: Regardless of angle and materials the rotational freedom increased in all groups. Volume loss with zirconia abutment was more than double than with a titanium abutment. Conclusion: The loose screw in implant prosthesis may lead to the wear of the hexagon on both components. This wear is sufficient to compromise the connection beyond repairpt_PT
dc.description.versioninfo:eu-repo/semantics/publishedVersionpt_PT
dc.identifier.citationJosé AR, et al. Wear of the implant connection with zirconia and titanium abutments. J Dental Sci 2017, 2(6): 000149. DOI 10.23880/oajds-16000149pt_PT
dc.identifier.doi10.23880/oajds-16000149pt_PT
dc.identifier.issn2573-8771
dc.identifier.urihttp://hdl.handle.net/10400.26/29323
dc.language.isoengpt_PT
dc.publisherMedwin Publisherspt_PT
dc.relation.publisherversionhttps://medwinpublishers.com/OAJDS/OAJDS16000149.pdfpt_PT
dc.subjectTitanium abutmentpt_PT
dc.subjectZirconia abutmentpt_PT
dc.subjectLoose screwpt_PT
dc.titleWear of the implant connection with zirconia and titanium abutmentspt_PT
dc.typejournal article
dspace.entity.typePublication
oaire.citation.startPage000149pt_PT
oaire.citation.titleJournal of Dental Sciencespt_PT
oaire.citation.volume2(6)pt_PT
rcaap.rightsopenAccesspt_PT
rcaap.typearticlept_PT

Ficheiros

Pacote original

A mostrar 1 - 1 de 1
A carregar...
Miniatura
Nome:
Artigo_PMaurício_2017_04.pdf
Tamanho:
1.03 MB
Formato:
Adobe Portable Document Format

Licença do pacote

A mostrar 1 - 1 de 1
A carregar...
Miniatura
Nome:
license.txt
Tamanho:
1.85 KB
Formato:
Item-specific license agreed upon to submission
Descrição: